October 19, 2026

Minimally invasive endoscopic surgery has revolutionized modern medicine. In Germany, for example, more than 6 million gastrointestinal endoscopies are performed each year. As a core functional component of endoscopes, the endoscope bending section is critical in clinical diagnosis and treatment. Currently deployed clinical endoscopes are complex in structure, costly, and the instruments are reused. Even after thorough cleaning and disinfection, the risk of cross-infection remains. To address this pain point, Single-Use Instruments (SUI) have become an industry trend. Single-use endoscopes impose requirements on functional components such as the endoscope bending section: simple structure, controllable cost, and ease of processing and assembly.

Commercially available endoscopes with an outer diameter of 10 mm use a conventional endoscope bending section, in which multiple cylindrical units are connected by rivets/cylindrical lugs to form a sliding hinge structure. The base material is mostly biocompatible stainless steel.

Conventional riveted endoscope bending sections have notable drawbacks: a large number of discrete parts and complex assembly procedures drive up manufacturing costs; riveting clearances introduce axial and rotational play, reducing steering control accuracy; and the mechanism lacks an inherent initial position, so the bending joint cannot return to its original position autonomously.

A monolithic nitinol flexure hinge structure can address these pain points. The integrated hinge requires no assembly, features zero friction, zero wear, and zero motion clearance, and can be directly fabricated in one step from a raw tube by laser tube cutting. The absence of secondary assembly makes it highly suitable for cost-sensitive single-use endoscopes.

Extensive research has been conducted worldwide on flexible bending hinges for endoscopes. Howell proposed a general design method for compliant mechanisms. Coemert et al. optimized the stress distribution of flexure hinges to improve fatigue life, but their approach was not adapted to nitinol superelastic materials. Existing literature has reported various nitinol flexible actuators: an electrical discharge machined segmented nitinol hinge for frontal sinus surgery, and a single-plane bending nitinol nested-tube manipulator. However, most of them support only single-plane deflection and cannot achieve the three-dimensional spherical motion capability required by modern flexible endoscopes (bending angle ≥ ±90°). Conventional biocompatible alloys such as stainless steel cannot withstand the large reversible deformation required by the hinge.

In nitinol alloys, the atomic ratio of nickel to titanium is nearly equal, giving them an austenite/martensite two-phase crystal structure, shape memory effect, and excellent superelasticity. It can achieve 6–8% reversible strain and is an ideal base material for single-use endoscope bending components. Adjusting the Ni/Ti ratio can tune the transformation temperature. The material's biocompatibility has been verified through extensive in vitro and in vivo experiments, and it is widely used in high-end medical implants and minimally invasive instruments. Various constitutive models for nitinol superelasticity have been established in academia and can be imported into ANSYS and ABAQUS for finite element simulation.

Nitinol has high hardness and thermal sensitivity; conventional mechanical machining can readily degrade the material's mechanical properties. Ultrashort-pulse laser cutting (picosecond/femtosecond lasers) of nitinol tubes produces a minimal heat-affected zone and fully preserves the material's microstructure and mechanical properties. This is the preferred process for manufacturing monolithic nitinol bending sections.

For endoscopic bending actuation, nitinol is often used as a shape memory actuation element. Nitinol wire, nitinol spring, and nitinol leaf spring approaches have been widely studied, but most of them suffer from complex assembly, high control difficulty, and the need for insulation/cooling accessories.

Monolithic nitinol bending sections are manufactured integrally by laser cutting nitinol tubes for single-use minimally invasive endoscopes. Compared with conventional stainless steel riveted endoscope bending sections, the required actuation force can be reduced by 50%, and the inner lumen diameter is increased by about 10%, reserving ample space for the cables and channels inside the instrument. The mechanism has an inherent home position for resetting and produces no axial clearance or friction loss during operation. During the design phase, the axial preload caused by the nitinol actuator must be fully evaluated. Limited by the cylindrical solid geometry, this monolithic approach cannot achieve the minimum bending radius of conventional riveted bending sections (conventional riveted bending sections themselves were not designed for nitinol actuation).

Contact Us